SYSTEM 1E Liquid Chemical Sterilant Processing System, Distributed by STERIS Corporation, Mentor, OH The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for the liquid chemical sterilization of manually cleaned immersible and reusable critical and semi-critical heat sensitive medical devices, including endoscopes and their accessories.

Recall: Z-0616-2014 · Initiated November 21, 2013

Recall

Recall Number
Z-0616-2014
Event Number
67002
Firm
Steris Corporation
FEI Number
1527821
Product Code
MLR
Status
Terminated
Root Cause
Device Design
Initiated
November 21, 2013
Posted
December 31, 2013
Terminated
June 16, 2014
Address
5960 Heisley Rd, Mentor, OH, 44060-1834

Description

SYSTEM 1E Liquid Chemical Sterilant Processing System, Distributed by STERIS Corporation, Mentor, OH The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for the liquid chemical sterilization of manually cleaned immersible and reusable critical and semi-critical heat sensitive medical devices, including endoscopes and their accessories.

Reason

Customers indicated fluctuations in environmental conditions sites and gradual degradation/deformation of check valve o-rings following frequent contact with S40 sterilant use solution result in nuisance aborts.

Action

STERIS sent an Urgent Voluntary Field Correction Notice dated November 21, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed that a STERIS Field Service Representative would contact their facility to schedule installation of the new hardware and updated software on their SYSTEM IE Liquid Chemical Sterilant Processing System(s). Customers were instructed that they may continue to operate the SYSTEM IE processor prior to the completion of the update. Customers with questions were instructed to contact STERIS Field Service Dispatch at 1-800-333-8828. For questions regarding this recall call 440-392-7519.

Distribution

Nationwide Distribution - USA (nationwide) and the District of Columbia and Puerto Rico. Product was also shipped to the following countries: Afghanistan, Bahrain, Hong Kong, Italy, Saudi Arabia, Turkey & United Kingdom.

Quantity

6,028 units