FDA Recall Terminated

COBAS AmpliPrep Instrument (CAP) with AMPLILINK (AL) Software v.3.3.5 Automated sample processing instrument used with CAP/CA and CAP/CTM. Roche Molecular Systems, Inc. 1080 US Highway 202 South Branchburg, NJ 08876 USA Made in Switzerland

Recall: Z-0614-2013 · Initiated October 11, 2012

Recall

Recall Number
Z-0614-2013
Event Number
63814
Firm
Roche Molecular Systems, Inc.
FEI Number
2243471
Product Code
JJH
Status
Terminated
Root Cause
Device Design
Initiated
October 11, 2012
Posted
December 23, 2012
Terminated
September 11, 2013
Address
1080 Us Highway 202 S, Branchburg, NJ, 08876-3733

Description

COBAS AmpliPrep Instrument (CAP) with AMPLILINK (AL) Software v.3.3.5 Automated sample processing instrument used with CAP/CA and CAP/CTM. Roche Molecular Systems, Inc. 1080 US Highway 202 South Branchburg, NJ 08876 USA Made in Switzerland

Reason

Certain COBAS AmpliPrep instruments using AMPILINK software v.3.3.4 had the sample clot detection flagging feature disabled during quality control release resulting in samples that may not be flagged as having clots.

Action

Roche sent an Urgent Medical Device Correction notification/Fax Back forms dated 10/11/12 to potentially affected customers. The letter identified the affected product, the issue, clinical significance, workaround, and required actions, along with contact information. Customers are to use the workaround until their Roche FSE has performed a Master Initialization on their affected instruments. Also, the letter should be filed for future reference. Questions should be directed to Roche Molecular Diagnostics Technical Support at 1-800-526-1247.

Distribution

Distributed in the states of GA, CA, NC, and MN.

Quantity

7 units