COBAS AmpliPrep Instrument (CAP) with AMPLILINK (AL) Software v.3.3.5 Automated sample processing instrument used with CAP/CA and CAP/CTM. Roche Molecular Systems, Inc. 1080 US Highway 202 South Branchburg, NJ 08876 USA Made in Switzerland
Recall
- Recall Number
- Z-0614-2013
- Event Number
- 63814
- Firm
- Roche Molecular Systems, Inc.
- FEI Number
- 2243471
- Product Code
- JJH
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- October 11, 2012
- Posted
- December 23, 2012
- Terminated
- September 11, 2013
- Address
- 1080 Us Highway 202 S, Branchburg, NJ, 08876-3733
Description
COBAS AmpliPrep Instrument (CAP) with AMPLILINK (AL) Software v.3.3.5 Automated sample processing instrument used with CAP/CA and CAP/CTM. Roche Molecular Systems, Inc. 1080 US Highway 202 South Branchburg, NJ 08876 USA Made in Switzerland
Certain COBAS AmpliPrep instruments using AMPILINK software v.3.3.4 had the sample clot detection flagging feature disabled during quality control release resulting in samples that may not be flagged as having clots.
Roche sent an Urgent Medical Device Correction notification/Fax Back forms dated 10/11/12 to potentially affected customers. The letter identified the affected product, the issue, clinical significance, workaround, and required actions, along with contact information. Customers are to use the workaround until their Roche FSE has performed a Master Initialization on their affected instruments. Also, the letter should be filed for future reference. Questions should be directed to Roche Molecular Diagnostics Technical Support at 1-800-526-1247.
Distributed in the states of GA, CA, NC, and MN.
7 units