FDA Recall Terminated

NexGen Complete Knee Solution Rotating Hinge Knee tibial broach impactor, Catalog # 00-5881-080-00, Zimmer Inc., Warsaw, IN.

Recall: Z-0610-2011 · Initiated April 16, 2010

Recall

Recall Number
Z-0610-2011
Event Number
55363
Firm
Zimmer Inc.
FEI Number
1000220733
Product Code
LXH
Status
Terminated
Root Cause
Device Design
Initiated
April 16, 2010
Posted
December 13, 2010
Terminated
September 12, 2012
Address
345 E Main St, Warsaw, IN, 46580-2746

Description

NexGen Complete Knee Solution Rotating Hinge Knee tibial broach impactor, Catalog # 00-5881-080-00, Zimmer Inc., Warsaw, IN.

Reason

The impactor may disassemble or detach during use. The detached piece may not be observed intraoperatively and may be left in the surgical site. Complications may include pain, infection, and prolongation in surgery. These complications may also require revision surgery.

Action

Zimmer issued "Urgent Medical Device Safety Alert" letters dated April 16, 2010 to Risk Managers, user facilities, Health Care Professionals and Zimmer Sales force. The users were informed to inspect the device for loosening. The alert also informed users to inspect the device after each use to verify that the device was intact. Customers were instructed to complete and return a User Facility Documentation Form. Zimmer will repair and replace all affected devices. Customers can contact ZImmer, Inc. at 800 613-6131 concerning this recall.

Distribution

Worldwide Distribution: Throughout US including Puerto Rico, and the countries of Argentina, Australia, Belgium, Bolivia, Brazil, Canada, Chile, China, Columbia, Denmark, Ecuador, El Salvador, France, Germany, Guatemala, Hong Kong, India, Italy, Japan, Korea, Mexico, Netherlands, Nicaragua, Panama, Peru, Portugal, Singapore, Spain, Switzerland, Taiwan, Thailand, Turkey, Ukraine, and United Kingdom.

Quantity

10,666 both products