FDA Recall Terminated

Carousel MAS1025-01, part of the Brain Surgical Accessories Intra-operative Kit (KIT0295), accessory used with the Renaissance System (TPL0038). Indicated for precise positioning of surgical instruments or spinal implants during general spinal and brain surgery. It is intended to be used whenever the clinician and/or patient benefits from generated 3D imaging of, high contrast objects.

Recall: Z-0609-2019 · Initiated November 26, 2014

Recall

Recall Number
Z-0609-2019
Event Number
81521
Firm
Mazor Robotics Ltd
FEI Number
3005075696
Product Code
HAW
Status
Terminated
Root Cause
Device Design
Initiated
November 26, 2014
Terminated
July 3, 2019
Address
5 Shacham Street P.O. Box 3104, North Industrial Park Israel

Description

Carousel MAS1025-01, part of the Brain Surgical Accessories Intra-operative Kit (KIT0295), accessory used with the Renaissance System (TPL0038). Indicated for precise positioning of surgical instruments or spinal implants during general spinal and brain surgery. It is intended to be used whenever the clinician and/or patient benefits from generated 3D imaging of, high contrast objects.

Reason

In some situations the carousel position pin (locker) may cause the carousel upper plate to pop out and the carousel may then inadvertently move.

Action

All Mazor Brain Clinical Sales Representatives (support during clinical cases) were notified of the issue on 11/26/2014 via email as well as received training.

Distribution

Affected devices were distributed in the US to four different medical facilities in four states

Quantity

5 units