Hitachi CXR4 Computed Tomography Scanner. This product is an x-ray imaging device that produces cross-sectional images of he body at different angles. The system reconstructs, processes, displays, and stores the collected images. The device output can provide an aid to diagnosis when used by a qualified physician and is intended for general purpose CT applications.
Recall
- Recall Number
- Z-0608-2009
- Event Number
- 50167
- Firm
- Hitachi Medical Systems America Inc
- FEI Number
- 1528028
- Product Code
- JAK
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- November 5, 2008
- Posted
- January 8, 2009
- Terminated
- September 21, 2012
- Address
- 1959 Summit Commerce Park, Twinsburg, OH, 44087
Description
Hitachi CXR4 Computed Tomography Scanner. This product is an x-ray imaging device that produces cross-sectional images of he body at different angles. The system reconstructs, processes, displays, and stores the collected images. The device output can provide an aid to diagnosis when used by a qualified physician and is intended for general purpose CT applications.
Software error: The device has a software error that can result in a condition where the displayed image orientation does not match the display orientation markers (right-left reversal).
On 11/5/2008 Hitachi Medical Systems sent recall notification letters entitled "URGENT - DEVICE RECALL CORRECTION" to its customers informing them of the problem with the software. They stated that a Hitachi service representative will contact each customer to schedule a software upgrade. For further information, please contact Hitachi Medical Systems America Inc. by telephone at 330-425-1313.
Nationwide Distribution --- including states of AZ, CA, DE, FL, ID, IL, IN, KS, KY, LA, MA, MO, NJ, NY, OH, OK, PA, PR, SC, TX, VA, and WI.