FDA Recall Terminated

SKYLight Imaging Systems, Model 882050; Intended to produce images depicting the anatomical distributions of single photon and positron emitting radioisotopes within the human body for interpretation by medical personnel.

Recall: Z-0607-2012 · Initiated December 8, 2011

Recall

Recall Number
Z-0607-2012
Event Number
60660
Firm
Philips Medical Systems
FEI Number
1525965
Product Code
KPS
Status
Terminated
Root Cause
Device Design
Initiated
December 8, 2011
Posted
January 11, 2012
Terminated
August 13, 2012
Address
3860 N 1st St, San Jose, CA, 95134-1702

Description

SKYLight Imaging Systems, Model 882050; Intended to produce images depicting the anatomical distributions of single photon and positron emitting radioisotopes within the human body for interpretation by medical personnel.

Reason

During preventative maintenance, a crack was discovered in the lower portion of the old style, non-common arm for the detector assembly, which could result in the detector dropping in an uncontrolled manner.

Action

Urgent Medical Device Correction/Field Safety Notices, dated Dec 08, 2011, were mailed to affected customer sites on December 9th and 12th, 2011 by Return Receipt certified mail. The letter identified the affected product, described the issue and when it could occur, along with the hazards involved. Customers were advised to stop using the system and contact a Philips service representative for an immediate inspection. Customers were also to follow the Instructions For Use to avoid the issue. Philips service will contact customers to arrange a time for Philips to inspect the system. The letter should be placed in the Instructions for Use. The Field Change Order will be released on December 16, 2011. If further information or support is needed, customers should contact Customer Care Solutions Center at 1-800-722-9377.

Distribution

Worldwide Distribution -- USA, Canada, Italy, Spain, France, United Kingdom, Sweden, Switzerland, Germany, Japan, Denmark, and the Netherlands.

Quantity

129