Liko AB Uno Lifts, Powered (non-AC) Patient Lift: Old Uno 102: S/N 20001 through 21300 New Uno 102 EM/EE/ES: S/N 30001 through 48100 Production Dates: Manufactured prior to June 26, 2008 Uno 100 and 102 are battery powered mobile lifts, designed to be used in most of the common lifting situations, for example transfers between bed and wheelchair, to and from the toilet and bathtub or for lifting to and from the floor.
Recall
- Recall Number
- Z-0606-2009
- Event Number
- 50155
- Firm
- Liko North America Inc
- FEI Number
- 3001451524
- Product Code
- FSA
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- November 11, 2008
- Posted
- January 27, 2009
- Terminated
- February 4, 2016
- Address
- 122 Grove St, Franklin, MA, 02038-3159
Description
Liko AB Uno Lifts, Powered (non-AC) Patient Lift: Old Uno 102: S/N 20001 through 21300 New Uno 102 EM/EE/ES: S/N 30001 through 48100 Production Dates: Manufactured prior to June 26, 2008 Uno 100 and 102 are battery powered mobile lifts, designed to be used in most of the common lifting situations, for example transfers between bed and wheelchair, to and from the toilet and bathtub or for lifting to and from the floor.
Actuator of the lift has the potential to separate when the upper arm is fully extended.
Hill Rom notified customers with a Field Safety Notice, dated December 5, 2008 to notify customers of the issue, potential risks, and to inspect units per their instructions and remove units from service if the unit fails inspection. Hill-Rom will provide a permanent fix, using an "OuterTube" to be placed on all units.
Worldwide Distribution --- including USA and country of Canada.
24070 total units