FDA Recall Terminated

VERIFY Dual Species Self-Contained Biological Indicator, Distributed by STERIS Corporation, Mentor, OH. Intended for use in installation testing and routine monitoring of steam sterilization and ethylene oxide sterilization processes.

Recall: Z-0605-2014 · Initiated September 10, 2013

Recall

Recall Number
Z-0605-2014
Event Number
67036
Firm
Steris Corporation
FEI Number
1527821
Product Code
FRC
Status
Terminated
Root Cause
Use error
Initiated
September 10, 2013
Posted
December 28, 2013
Terminated
February 14, 2014
Address
5960 Heisley Rd, Mentor, OH, 44060-1834

Description

VERIFY Dual Species Self-Contained Biological Indicator, Distributed by STERIS Corporation, Mentor, OH. Intended for use in installation testing and routine monitoring of steam sterilization and ethylene oxide sterilization processes.

Reason

STERIS has identified that the population of the G. stearothermophilus spores that monitor STEAM sterilization cycles does not meet specifications for Lot # 131004 of the Verify Dual Species Self-Contained Biological Indicators. As a result the spore count on the affected product is lower than the labeled minimum and the product may incorrectly indicate that a 5-log spore kill was achieved.

Action

On September 10, 2013, the firm sent Urgent Voluntary Recall Notices to their customers.

Distribution

Worldwide Distribution-USA (nationwide) including the states of AK, Al, AR, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and WY, and the countries of BAHRAIN, CANADA, COLUMBIA, GERMANY, ITALY, MALAYSIA, PHILIPPINES, ROMANIA, SPAIN, and THE NETHERLANDS.

Quantity

1,238 boxes