FDA Recall Terminated

Liko AB Uno Lifts, Powered (non-AC) Patient Lift: Uno 100 EM/EE: S/N 7090001 through 7096200 Uno 101: S/N 10001 through 11000 Production Dates: Manufactured prior to June 26, 2008 Uno 100 and 102 are battery powered mobile lifts, designed to be used in most of the common lifting situations, for example transfers between bed and wheelchair, to and from the toilet and bathtub or for lifting to and from the floor.

Recall: Z-0605-2009 · Initiated November 11, 2008

Recall

Recall Number
Z-0605-2009
Event Number
50155
Firm
Liko North America Inc
FEI Number
3001451524
Product Code
FSA
Status
Terminated
Root Cause
Process control
Initiated
November 11, 2008
Posted
January 27, 2009
Terminated
February 4, 2016
Address
122 Grove St, Franklin, MA, 02038-3159

Description

Liko AB Uno Lifts, Powered (non-AC) Patient Lift: Uno 100 EM/EE: S/N 7090001 through 7096200 Uno 101: S/N 10001 through 11000 Production Dates: Manufactured prior to June 26, 2008 Uno 100 and 102 are battery powered mobile lifts, designed to be used in most of the common lifting situations, for example transfers between bed and wheelchair, to and from the toilet and bathtub or for lifting to and from the floor.

Reason

Actuator of the lift has the potential to separate when the upper arm is fully extended.

Action

Hill Rom notified customers with a Field Safety Notice, dated December 5, 2008 to notify customers of the issue, potential risks, and to inspect units per their instructions and remove units from service if the unit fails inspection. Hill-Rom will provide a permanent fix, using an "OuterTube" to be placed on all units.

Distribution

Worldwide Distribution --- including USA and country of Canada.

Quantity

24070 total units