FDA Recall Open, Classified

GE Healthcare MR superconducting magnets, a component of GE Healthcare SIGNA 1.5T TwinSpeed MR System, nuclear magnetic resonance system.

Recall: Z-0604-2022 · Initiated December 23, 2021

Recall

Recall Number
Z-0604-2022
Event Number
89376
Firm
GE Healthcare, LLC
FEI Number
2126677
Product Code
LNH
Status
Open, Classified
Root Cause
Process control
Initiated
December 23, 2021
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

GE Healthcare MR superconducting magnets, a component of GE Healthcare SIGNA 1.5T TwinSpeed MR System, nuclear magnetic resonance system.

Reason

The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.

Action

The recalling firm issued letters dated 12/23/2021 via FedEx explaining the safety issue and the actions to be taken by the customer/user. The actions say that the customer can continue to use the system as normal but they are to follow the instructions listed in the letter and to ensure the system is inspected either by GE Healthcare or the organization who installed their ventilation system to confirm it complies with the safety requirements for ventilation. The customer was requested to complete the workflow which was to be accessed through a QR code or link in the letter within the Medical Device Notification Acknowledgement Response Required form. It was to be returned no later than 30 days from receipt. The customer is also told it is important they continue to follow the guidelines outlined in the Safety chapter of their system Operator Manual, including ensuring a procedure is in place to evacuate the patient and personnel from the magnet room should a quench occur.

Distribution

Distribution was nationwide, including Puerto Rico and the Virgin Islands. There was also government and military distribution. Foreign distribution was made to Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaidjan, Bahrain, Bangladesh, Belarus, Belgium, Bhutan, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Republic of Congo, Croatia, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, Equatorial Guinea, Estonia, Ethiopia, Finland, France, Georgia, Germany, Ghana, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Kyrhyzstan, Latvia, Lebanon, Libyan Arabic Jamahiriya, Lithuania, Luxembourg, Macedonia, Madagascar, Malaysia, Mali, Malta, Mauritania, Mexico, Republic of Moldova, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Nordic-HQ, Norway, Oman, Pakistan, State of Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Swaziland, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Vietnam, Zambia, and Zimbabwe.

Quantity

17,228 devices total in this field correction