FDA Recall Terminated

Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required) base; Model 801764. (Not distributed within the United States).

Recall: Z-0603-2007 · Initiated August 1, 2005

Recall

Recall Number
Z-0603-2007
Event Number
37049
Firm
Terumo Cardiovascular Systems Corp
FEI Number
1828100
Product Code
DTQ
Status
Terminated
Root Cause
Other
Initiated
August 1, 2005
Posted
June 30, 2007
Terminated
August 4, 2009
Address
6200 Jackson Rd, Ann Arbor, MI, 48103-9586

Description

Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required) base; Model 801764. (Not distributed within the United States).

Reason

The power supply may fail to function due to a defective capacitor. (Note; this device incorporates a redundant separate power supply not affected by this capacitor)

Action

The power supplies were replaced as complaints were received between August 2005 and February 2006. The firm issued a Device Correction Letter dated 6/6/07 notifying all users of the problem and advising them that their units will be corrected when replacement components become available.

Distribution

Worldwide Distribution: USA, Australia, Belgium, Canada, Dominican Republic, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Jordan, Malaysia, Mexico, Pakistan, Philippines, Russia, Saudi Arabia, Singapore and South Africa.

Quantity

534 of both model numbers