Artis Q, Angiographic x-ray system Product Usage: The Artis systems are a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis zee / zeego and Q/ Q.zen include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. Artis zee / zeego and Q/ Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis.
Recall
- Recall Number
- Z-0601-2017
- Event Number
- 75606
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2240869
- Product Code
- OWB
- Status
- Terminated
- Root Cause
- Software in the Use Environment
- Initiated
- November 1, 2016
- Posted
- November 17, 2016
- Terminated
- June 21, 2017
- Address
- 40 Liberty Blvd, Malvern, PA, 19355-1418
Description
Artis Q, Angiographic x-ray system Product Usage: The Artis systems are a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis zee / zeego and Q/ Q.zen include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. Artis zee / zeego and Q/ Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis.
The possibility exists that the semi-automatic focus switch may not work as intended when pressing the footswitch multiple times.
Siemens mailed a Customer Safety Advisory Notice to affected customers to inform them of a corrective action they plan to perform to prevent a possible hazard to persons and equipment. The notice identified the associated risks, underlying issues, impact on system operation and actions to be taken.
US Nationwide Distribution.
4,657 systems (worldwide)