Software version VD10E for Syngo X-Workplace; Picture archiving and communication system.
Recall
- Recall Number
- Z-0597-2016
- Event Number
- 72751
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2240869
- Product Code
- LLZ
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- November 18, 2015
- Posted
- January 6, 2016
- Terminated
- December 5, 2016
- Address
- 40 Liberty Blvd, Malvern, PA, 19355-1418
Description
Software version VD10E for Syngo X-Workplace; Picture archiving and communication system.
Potential post-processing software issue when using tabcard "4D" on X-Workplace with software version VD10E. A too small measurement in the MPRs of a volume when images are acquired with a CT scanner with a tilted gantry. In a 2x2 layout, the 4D tabcard will show an incorrect, too small length measurement. This could result in selecting a device of the wrong size, which then needs to be exchanged.
Siemens sent an Important Customer Safety Notice dated November 17, 2015. The Notice identified the product, the problem, and the action to be taken by the customer. The potential issue will be remedied with a software update via Update Instructions AX067/15/S. Siemens service organization will contact customers to arrange a date for the installation of the software update. Customers with questions were advised to call 800-888-7436.
Nationwide Distribution including: AL, AZ, CA, CO, FL, IA, ID, IL, IN, MA, MD, MI, MN, MO, MT, NC, NY, OH, OK, PA, SC, TN, TX, VA, and WI.
73 units