FDA Recall Terminated

Software version VD10E for Syngo X-Workplace; Picture archiving and communication system.

Recall: Z-0597-2016 · Initiated November 18, 2015

Recall

Recall Number
Z-0597-2016
Event Number
72751
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
LLZ
Status
Terminated
Root Cause
Software design
Initiated
November 18, 2015
Posted
January 6, 2016
Terminated
December 5, 2016
Address
40 Liberty Blvd, Malvern, PA, 19355-1418

Description

Software version VD10E for Syngo X-Workplace; Picture archiving and communication system.

Reason

Potential post-processing software issue when using tabcard "4D" on X-Workplace with software version VD10E. A too small measurement in the MPRs of a volume when images are acquired with a CT scanner with a tilted gantry. In a 2x2 layout, the 4D tabcard will show an incorrect, too small length measurement. This could result in selecting a device of the wrong size, which then needs to be exchanged.

Action

Siemens sent an Important Customer Safety Notice dated November 17, 2015. The Notice identified the product, the problem, and the action to be taken by the customer. The potential issue will be remedied with a software update via Update Instructions AX067/15/S. Siemens service organization will contact customers to arrange a date for the installation of the software update. Customers with questions were advised to call 800-888-7436.

Distribution

Nationwide Distribution including: AL, AZ, CA, CO, FL, IA, ID, IL, IN, MA, MD, MI, MN, MO, MT, NC, NY, OH, OK, PA, SC, TN, TX, VA, and WI.

Quantity

73 units