FDA Recall Open, Classified

CARDIOHELP-i Heart Lung Support System: Blood oxygenation and carbon dioxide removal system used to pump blood through the extracorporeal bypass circuit for circulatory and/or pulmonary support for periods appropriate to cardiopulmonary bypass (up to six hours). Product REF Numbers: 701072780; 701048012

Recall: Z-0595-2024 · Initiated November 14, 2023

Recall

Recall Number
Z-0595-2024
Event Number
93462
Firm
Maquet Medical Systems USA
FEI Number
3008355164
Product Code
DTQ
Status
Open, Classified
Root Cause
Process design
Initiated
November 14, 2023
Posted
September 30, 2024
Address
45 Barbour Pond Dr, Wayne, NJ, 07470-2094

Description

CARDIOHELP-i Heart Lung Support System: Blood oxygenation and carbon dioxide removal system used to pump blood through the extracorporeal bypass circuit for circulatory and/or pulmonary support for periods appropriate to cardiopulmonary bypass (up to six hours). Product REF Numbers: 701072780; 701048012

Reason

CARDIOHELP-i System was not properly tested to measure leakage current

Action

Maquet Cardiopulmonary GmbH (MCP)/Getinge issued Urgent Medical Device Correction Letter to U.S, consignees November 14, 2023 via FedEx. Letter states reason for recall, health risk and action to take: Please examine your inventory immediately to determine if you have any affected CARDIOHELP-i in your inventory. " Whether or not your facility has affected product(s) listed in this notice, please complete and sign the attached MEDICAL DEVICE - CORRECTION RESPONSE FORM (Page 4) to acknowledge that you have received this notification. Return the completed form to Getinge by e-mailing a scanned copy to [email protected] or by faxing the form to 1-800-530-6084. " If you report having any affected CARDIOHELP-i devices, a local Getinge representative will contact you to arrange a system check of the device to ensure that leakage current is within specifications. " Affected CARDIOHELP-i devices are not requested to be returned and can be used as usual. " Please forward this information to all current and potential CARDIOHELP-i users within your hospital / facility. " If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. Type of Action by Getinge: Getinge has identified the issue to be related to a supplier using an incorrect drawing to produce a tool used to measure leakage current of the CARDIOHELP-i. This issue has been corrected by the supplier. MCP/Getinge will contact you to arrange for a MCP/Getinge service technician to perform a system check of any affected device you report to confirm leakage current is within specifications (IEC 62353). If a system fails the performed test, the MCP/Getinge service technician will replace the system s sensor panel for you. MCP/Getinge service may have already performed the required system check of your device during a past preventative maintenance. MCP/Getinge will check service records to confirm. MCP

Distribution

Nationwide-WW Distribution to Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, French Polynesia, Germany, Ghana, Greece, Hong Kong, Iceland, India, Iran, Islamic Republic Of, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Republic Of Kuwait, Lebanon, Luxembourg, Macao, Maldives, Martinique, Mayotte, Mexico, Netherlands, New Zealand, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom,Vietnam, Yemen

Quantity

1594 units