FDA Recall Terminated

Vysis LSI ATM/p53: D13S319 /CEP 12/13q34 DNA Probe Set; fluorescence in situ hybridization (FISH) analyte specific reagents, in vitro diagnostic; Abbott Molecular Inc., Des Plaines, IL 60018 USA; list 05J83-001; The probe set consists of a 200 microliter vial of Vysis LSI D13S319 SO/LSI 13q34 SA/CEP 12 SG Probe, reference 30-191024, and a 200 microliter vial of Vysis LSI ATM SG/p 53 SO Probe, reference 30-191025.

Recall: Z-0595-2007 · Initiated December 22, 2006

Recall

Recall Number
Z-0595-2007
Event Number
37363
Firm
Abbott Molecular
FEI Number
3005248192
Product Code
MVU
Status
Terminated
Root Cause
Other
Initiated
December 22, 2006
Posted
March 13, 2007
Terminated
April 17, 2008
Address
1300 E Touhy Ave, Des Plaines, IL, 60018-3315

Description

Vysis LSI ATM/p53: D13S319 /CEP 12/13q34 DNA Probe Set; fluorescence in situ hybridization (FISH) analyte specific reagents, in vitro diagnostic; Abbott Molecular Inc., Des Plaines, IL 60018 USA; list 05J83-001; The probe set consists of a 200 microliter vial of Vysis LSI D13S319 SO/LSI 13q34 SA/CEP 12 SG Probe, reference 30-191024, and a 200 microliter vial of Vysis LSI ATM SG/p 53 SO Probe, reference 30-191025.

Reason

The LSI D13S319 SO/13q34 SA/CEP 12 SG probe of the two mixture DNA-Probe was contaminated with LSI ATM SG/p53 SO probe. This contamination would result in 10 FISH signals in each normal cell rather than the expected 6 FISH signals in each normal cell.

Action

Abbott Molecular telephoned and/or sent recall letters to the Abbott FISH accounts on 12/22/06, advising them that an unexpected signal pattern with the Vysis LSI ATM/p53: D13S319 /CEP 12/13q34 Probe was observed, confirmed and identified as a result of contamination in a subset of vials of D13S319 /CEP 12/13q34 from lot number 79500. This contamination results in 10 FISH signals in each normal cell (4 SO, 4 SG, 2 SA) rather than the expected 6 FISH signals in each normal cell (2 SO, 2 SG, 2 SA). The customers were advised that lot 80178 was also being included in the recall, and were requested to review the information carefully and assess the risk to their patient population; discontinue and destroy any unused product immediately; complete and return the enclosed reply form; review the information with laboratory personnel; and communicate the information to health care providers served in cases where the affected product were used and reported. Abbott Molecular provided replacement product for that product destroyed, upon receipt of the reply form. Any questions were directed to Abbott Molecular Technical Service at 800-553-7042, option 2.

Distribution

Worldwide including USA, Austria, Canada, Czech Republic, Denmark, Finland, France, Germany, Italy, Netherlands, Norway, Spain, and the United Kingdom.

Quantity

109 probe sets