CareLink Clinic, REF: MMT-7350
Recall
- Recall Number
- Z-0594-2026
- Event Number
- 97873
- Firm
- Medtronic MiniMed, Inc.
- FEI Number
- 3003166194
- Product Code
- PHV
- Status
- Open, Classified
- Root Cause
- Software change control
- Initiated
- October 21, 2025
- Posted
- November 25, 2025
- Address
- 18000 Devonshire St, Northridge, CA, 91325-1219
Description
CareLink Clinic, REF: MMT-7350
Software error causing incorrect data to be displayed on the 24-hour Sensor Glucose Overview Graph, any potential therapy decisions were made based on the incorrect data displayed on the 24-hour sensor glucose overview graph may lead to hypoglycemia or hyperglycemia.
On 10/21/2025, "URGENT MEDICAL DEVICE CORRECTION" notices were emailed to customers informing them of the following: software version 4.2C was released to correct the issues with the data displayed on 24-hour Sensor Glucose Overview graph. If you made individual diabetes treatment recommendations to patients based on the hypoglycemic patterns or on the 30-day sensor glucose data displayed on the 24-hour Sensor Glucose Overview graph with affected software, revisit those recommendations utilizing patients individual CareLink reports, as they were not impacted by this issue. Complete and return the confirmation form: https://info.medtronicdiabetes.com/carelink-clinic-dashboard-hcp or email form to [email protected] For questions, contact the firm's 24-Hour Technical Support line at 1-800-646-4633, option 9.
Worldwide Distribution: US (nationwide) including states of: TX, NC, MN, CA, VA, ID, WA, GA, MI, MA, OH, IL, IA, CT, FL, PA, NY, TN, AZ, LA, AL, NJ, CO, SC, MS, OR, MT, MD, KS, AK, NH, NE, WY, ND, IN, MO, WI, KY, UT, NV, SD, WV, NM, HI, AR, OK, VT, RI; and OUS (foreign) to countries of: Australia (AU), Sweden (SE), Belgium (BE), Great Britain (GB), Netherlands (NL), Finland (FI), Canada (CA), Germany (DE), Spain (ES), Switzerland (CH), Italy (IT), South Africa (ZA), South Korea (KR), Singapore (SG), New Zealand (NZ), Poland (PL), Israel (IL), Denmark (DK), Norway (NO), Luxembourg (LU), Lithuania (LT), Czech Republic ( CZ), Mexico (MX), Brazil (BR), Colombia (CO), Slovakia (SK), Ireland (IE), Argentina (AR), Greece (GR), Kuwait (KW), Portugal (PT), Iceland (IS), Japan (JP), Turkey (TR), Austria (AT), Qatar (QA), Hong Kong (HK), Chile (CL), Ukraine (UA), Hungary (HU), Estonia (EE), RS, Latvia (LV), Taiwan (TW), Saudia Arabia (SA), Bahrain (BH), Oman (OM), Romania (RO), India (IN), Uruguay (UY), Malaysia (MY), United Arab Emirates (AE), Algeria (DZ), Croatia (HR), France (FR), Russia (RU), Cyprus (CY), Thailand (TH), Egypt (EG), Bosnia and Herzegovina (BA), Libya (LY), and Iraq (IQ)
12126