FDA Recall Terminated

Hoana LifeBed Patient Vigilance System, (Intelligent Medical Vigilance System),Model Number: 010135-01, Product distributed by Hoana Medical, Inc., 828 Fort St. Mall, Suite 620, Honolulu, HI

Recall: Z-0592-2008 · Initiated November 7, 2007

Recall

Recall Number
Z-0592-2008
Event Number
46154
Firm
Hoana Medical
FEI Number
3005596882
Product Code
KMI
Status
Terminated
Root Cause
Software design
Initiated
November 7, 2007
Posted
January 24, 2008
Terminated
February 1, 2008
Address
828 Fort Street Mall, Ste 620, Honolulu, HI, 96813-4314

Description

Hoana LifeBed Patient Vigilance System, (Intelligent Medical Vigilance System),Model Number: 010135-01, Product distributed by Hoana Medical, Inc., 828 Fort St. Mall, Suite 620, Honolulu, HI

Reason

Bed Exit Failure: Failure to recognize "Bed Exit" when the feature is active due to a software anomaly.

Action

On 11/7/07, all consignees were notified via an untitled letter and on-site visit by the firm's sales representative, informing them of the affected product and providing instructions on the recall. The recalling firm will arrange for installation of a software fix at each consignee.

Distribution

Nationwide including states of HI, CA, TX and MA

Quantity

216 units