FDA Recall Terminated

Maquet Servo Ventilator SV900C/D/E (SV900) Siemens-Elema AB, Solna, Sweden

Recall: Z-0589-2010 · Initiated November 17, 2009

Recall

Recall Number
Z-0589-2010
Event Number
53843
Firm
MAQUET Inc.
FEI Number
3008355164
Product Code
CBK
Status
Terminated
Root Cause
Device Design
Initiated
November 17, 2009
Posted
January 5, 2010
Terminated
June 4, 2012
Address
45 Barbour Pond Drive, Wayne, NJ, 07470

Description

Maquet Servo Ventilator SV900C/D/E (SV900) Siemens-Elema AB, Solna, Sweden

Reason

Some Servo Ventilators 300/300A and Servo Ventilator 900C/D/E should not be used with a system that may generate negative pressure below -100 cm H2O (Closed system suctioning) due to the risk of damaging the ventilator's pressure transducers.

Action

Maquet sent Device Correction letters commencing 11/17/09 and ending 11/23/09 by Federal Express.

Distribution

Worldwide distribution

Quantity

29364 (8438 of those in US)