FDA Recall Terminated

DIST LAT FEM LOCK PLT, RT

Recall: Z-0588-2014 · Initiated November 5, 2013

Recall

Recall Number
Z-0588-2014
Event Number
66499
Firm
Zimmer, Inc.
FEI Number
1000220733
Product Code
HRS
Status
Terminated
Root Cause
Process change control
Initiated
November 5, 2013
Posted
December 27, 2013
Terminated
May 5, 2015
Address
345 E Main St, Warsaw, IN, 46580-2746

Description

DIST LAT FEM LOCK PLT, RT

Reason

Based on a review of historic packaging validations, it was determined that a subset of product packaged at a Zimmer facility required testing to ensure that previously packaged product from this facility met the established requirements for sterile barrier integrity. This testing determined that a packaging tray configuration which consists of an inner and outer sterile tray, resulted in failure

Action

On 11/5/2013, Urgent Medical Device Recall Notifications were sent to affected distributors, hospital Risk Managers, and doctors informing them of the failure. All distributors were notified via electronic mail. Hospital risk managers, surgeons and distributors with affected inventory within the expiration date of the device will also be notified via courier. The letter identifies the issue, health risks, and their responsibilities. Questions concerning the recall are directed to 1-877-946-2761 Communications outside of the United States will occur approximately two weeks after the United States communications. Zimmer will conduct effectiveness checks in the following way: Distributors/Hospitals/Surgeons 100% of the notifications will be tracked to ensure delivery of the notifications. An inventory certification form will need to be returned from each distributor that certifies that all available affected product from their territory has been returned. Additional notifications via email and FedEx will be sent to those who have not completed the required certification form or returned affected product. Accounts will be deemed unresponsive after 3 attempts.

Distribution

Worldwide Distribution.

Quantity

42, 065 units total