FDA Recall
Terminated
Dimension LIP Flex Reagent Cartridge, Catalog# DF55A
Recall: Z-0588-2011
·
Initiated March 10, 2010
Recall
- Recall Number
- Z-0588-2011
- Event Number
- 55571
- Firm
- Siemens Healthcare Diagnostics, Inc.
- FEI Number
- 2517506
- Product Code
- CHI
- Status
- Terminated
- Root Cause
- Other
- Initiated
- March 10, 2010
- Posted
- December 10, 2010
- Terminated
- August 4, 2011
- Address
- 500 GBC Drive PO BOX 6101, Mailstop 514, Newark, DE, 19714-6101
Description
Dimension LIP Flex Reagent Cartridge, Catalog# DF55A
Reason
Open Well instability: After calibration, QC shifts high (out of range) after the Flex Well is hydrated. Shifts of up to 20% have been observed in both QC and patient samplesas soon as two hours post hydration.
Action
Recalling firm issued "Urgent Field Safety Notice", dated March 2010, informing users of the problem and to transition to the revised Lipasae method.
Distribution
Nationwide Distribution throughout the US.
Quantity
21171