FDA Recall
Terminated
Maquet Servo Ventilator 300/300A (SV300) Siemens - Elema AB,
Recall: Z-0588-2010
·
Initiated November 17, 2009
Recall
- Recall Number
- Z-0588-2010
- Event Number
- 53843
- Firm
- MAQUET Inc.
- FEI Number
- 3008355164
- Product Code
- CBK
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- November 17, 2009
- Posted
- January 5, 2010
- Terminated
- June 4, 2012
- Address
- 45 Barbour Pond Drive, Wayne, NJ, 07470
Description
Maquet Servo Ventilator 300/300A (SV300) Siemens - Elema AB,
Reason
Some Servo Ventilators 300/300A and Servo Ventilator 900C/D/E should not be used with a system that may generate negative pressure below -100 cm H2O (Closed system suctioning) due to the risk of damaging the ventilator's pressure transducers.
Action
Maquet sent Device Correction letters commencing 11/17/09 and ending 11/23/09 by Federal Express.
Distribution
Worldwide distribution
Quantity
20837 (7847 of those in US)