FDA Recall Terminated

CrossCath Support Catheter, RPN CXC3.0-1.9-14-90-P-NS-0, Order Number G51572 For diagnostic and interventional vascular procedures.

Recall: Z-0587-2020 · Initiated November 8, 2019

Recall

Recall Number
Z-0587-2020
Event Number
84212
Firm
Cook Inc.
FEI Number
1820334
Product Code
KRA
Status
Terminated
Root Cause
Process control
Initiated
November 8, 2019
Posted
December 9, 2019
Terminated
February 1, 2021
Address
750 N Daniels Way, Bloomington, IN, 47404-9120

Description

CrossCath Support Catheter, RPN CXC3.0-1.9-14-90-P-NS-0, Order Number G51572 For diagnostic and interventional vascular procedures.

Reason

A manufacturing error may cause the radiopaque marker bands to be too loose on certain CXC3.0 CrossCath Support Catheters (compatible with 0.014 wire guides) and too tight on certain CXC3.4 CrossCath Support Catheters (compatible with 0.018 wire guides).

Action

Urgent Medical Device Recall notification letters dated 11/8/19 were sent. Actions to be Taken by the Customer 1. Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all distribution and use of this product. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. Note: Unaffected products that are returned will not be credited. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return via fax (812.339.7316) or email ([email protected]). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: [email protected].

Distribution

Distribution to US states of AL, CA, CT, FL, GA, IL, MA, MD, MI, MO, NC, NJ, NY, OH, OK, PA, SC, TN, TX, and VA, and Brazil.

Quantity

20