FDA Recall Terminated

Implantable Cardioverter Defibrillators (ICDs), Evera. Labeled as the following: a. Evera S DR ICD DF1 (Product No. DDBC3D1); b. Evera S VR ICD DF1 (Product No. DVBC3D1); c. Evera S VR ICD DF4 (Product No. DVBC3D4); d. Evera XT DR ICD DF1 (Product No. DDBB1D1); e. Evera XT VR ICD DF1 (Product No. DVBB1D1); f. Evera XT VR ICD DF4 (Product No. DVBB1D4); EXPANSION: g. Evera S DR ICD DF4 (Product No. DDBC3D4); h. Evera XT DR ICD DF1 Gold (Product No. DDBB1D1G); i. Evera XT DR ICD DF4 (Product No. DDBB1D4) Implantable devices that automatically detect and treat episodes of ventricular fibrillation, ventricular tachycardia, fast ventricular tachycardia, and bradyarrhythmia. Some devices can also provide resynchronization therapy, including sequential biventricular pacing (CRT-D devices).

Recall: Z-0587-2018 · Initiated December 19, 2017

Recall

Recall Number
Z-0587-2018
Event Number
78888
Firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
FEI Number
2182208
Product Code
LWS
Status
Terminated
Root Cause
Process design
Initiated
December 19, 2017
Posted
February 1, 2018
Terminated
August 27, 2021
Address
8200 Coral Sea St Ne, Mounds View, MN, 55112-4391

Description

Implantable Cardioverter Defibrillators (ICDs), Evera. Labeled as the following: a. Evera S DR ICD DF1 (Product No. DDBC3D1); b. Evera S VR ICD DF1 (Product No. DVBC3D1); c. Evera S VR ICD DF4 (Product No. DVBC3D4); d. Evera XT DR ICD DF1 (Product No. DDBB1D1); e. Evera XT VR ICD DF1 (Product No. DVBB1D1); f. Evera XT VR ICD DF4 (Product No. DVBB1D4); EXPANSION: g. Evera S DR ICD DF4 (Product No. DDBC3D4); h. Evera XT DR ICD DF1 Gold (Product No. DDBB1D1G); i. Evera XT DR ICD DF4 (Product No. DDBB1D4) Implantable devices that automatically detect and treat episodes of ventricular fibrillation, ventricular tachycardia, fast ventricular tachycardia, and bradyarrhythmia. Some devices can also provide resynchronization therapy, including sequential biventricular pacing (CRT-D devices).

Reason

Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).

Action

On 12/19/2017, Medtronic verbally notified consignees and retrieved eight devices at sites that had affected inventory on hand. On 01/22/2018, Medtronic provided a recall notification letter to physicians following 48 patients implanted with affected devices and risk managers of those medical facilities, recommending that physicians strongly consider prophylactic device replacement for patients implanted with an affected device. Medtronic Technical Services is available to assist physicians with questions at 800-723-4636. Medtronic Patient Service is available to assist patients at 800-551-5544 (Monday-Friday, 8am-5pm Central Time). The recall was expanded on 03/15/2018 and customers were notified by letter. Affected product, not yet implanted, was retrieved on about 03/08/2018. Instructions in the physicians following patients implanted with the additional affected devices included prophylactic device replacement for those patients at higher risk, including patients whose clinical history indicates prior need for high-voltage therapy and/or for pacemaker-dependent patients. The physicians should carefully weigh the risks and benefits of device replacement. If it is determined that replacement is not warranted, instructions included the following: programming changes to reduce the potential for high-voltage charges associated with arrhythmia detection and therapies, such as enabling ATP before charging for fast ventricular rhythms or programming a separate fast VT via VF zone with ATP; Continue three-month in-clinic or remote follow-ups to verify device functionality because inability to interrogate a device or a failed remote monitoring transmission may be an indication that internal arcing has occurred; and to advice patients to seek medical attention immediately if they experience new or unexpected symptoms suspicious for a ventricular arrhythmia. Medtronic Patient Services is available to assist patients at 800-551-5544 (Monday-Friday, 8am-5pm Centra

Distribution

US Nationwide Distribution.

Quantity

951 devices total