MOVES System Lithium Polymer Batteries The MOVES System is a portable computer controlled electrically powered transport emergency ventilator intended to provide continuous or intermittent ventilatory support for the care of adults who require ventilation.
Recall
- Recall Number
- Z-0587-2016
- Event Number
- 72867
- Firm
- Thornhill Research Inc
- FEI Number
- 3005803793
- Product Code
- BTL
- Status
- Terminated
- Root Cause
- Component design/selection
- Initiated
- November 19, 2015
- Posted
- January 11, 2016
- Terminated
- April 12, 2019
- Address
- 210 Dundas St W, Toronto Canada Suite 200
Description
MOVES System Lithium Polymer Batteries The MOVES System is a portable computer controlled electrically powered transport emergency ventilator intended to provide continuous or intermittent ventilatory support for the care of adults who require ventilation.
Batteries found unresponsive after storage, not recognized by main device, and are perpetually enabled.
Thornhill Research sent a Recall Notification letter dated December 15, 2015 to affected customers. The letter identified the affected products, problem and actions to be taken. Customers were instructed to update their device software. For questions customers were instructed to call 416-597-1325.
Worldwide Distribution: US distribution in state of: Virginia and to the country of: Singapore.
178 units