FDA Recall Terminated

MOVES System, an emergency ventilator with suction, oxygen concentrator and multi-parameter patient monitoring capabilities. The MOVES System is a portable computer controlled electrically powered transport emergency ventilator intended to provide continuous or intermittent ventilatory support for the care of adults who require ventilation.

Recall: Z-0586-2016 · Initiated November 19, 2015

Recall

Recall Number
Z-0586-2016
Event Number
72867
Firm
Thornhill Research Inc
FEI Number
3005803793
Product Code
BTL
Status
Terminated
Root Cause
Component design/selection
Initiated
November 19, 2015
Posted
January 11, 2016
Terminated
April 12, 2019
Address
210 Dundas St W, Toronto Canada Suite 200

Description

MOVES System, an emergency ventilator with suction, oxygen concentrator and multi-parameter patient monitoring capabilities. The MOVES System is a portable computer controlled electrically powered transport emergency ventilator intended to provide continuous or intermittent ventilatory support for the care of adults who require ventilation.

Reason

Batteries found unresponsive after storage, not recognized by main device, and are perpetually enabled.

Action

Thornhill Research sent a Recall Notification letter dated December 15, 2015 to affected customers. The letter identified the affected products, problem and actions to be taken. Customers were instructed to update their device software. For questions customers were instructed to call 416-597-1325.

Distribution

Worldwide Distribution: US distribution in state of: Virginia and to the country of: Singapore.

Quantity

83 units