FDA Recall Terminated

Active I-PTH ELISA, Part Number: DSL-10-8000 The DSL-10-80000 ACTIVE I-PTH Enzyme-Linked Immunosorbent Assay (ELISA) Kit provides materials for the quantitative measurement of I-PTH in serum or plasma. This assay is intended for in vitro diagnostic use as an aid in the diagnosis of hyperparathyroidism.

Recall: Z-0584-2011 · Initiated April 22, 2010

Recall

Recall Number
Z-0584-2011
Event Number
56250
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
CEW
Status
Terminated
Root Cause
Pending
Initiated
April 22, 2010
Posted
December 9, 2010
Terminated
May 7, 2012
Address
200 S Kraemer Blvd, Brea, CA, 92821-6208

Description

Active I-PTH ELISA, Part Number: DSL-10-8000 The DSL-10-80000 ACTIVE I-PTH Enzyme-Linked Immunosorbent Assay (ELISA) Kit provides materials for the quantitative measurement of I-PTH in serum or plasma. This assay is intended for in vitro diagnostic use as an aid in the diagnosis of hyperparathyroidism.

Reason

The recall was initiated because Beckman Coulter has confirmed that the stated stability of the DSL-10-8000 I-PTH ELISA and DSL-8000 I-PTH IRMA assays may not be achieved for the lot numbers affected by the recall. The affected I-PTH ELISA and the I-PTH IRMA kit lots may produce falsely elevated patient results. Internal stability testing showed an average 16% elevation for I-PTH ELISA lot 99

Action

Beckman Coulter sent an Urgent Product Corrective Action Letter to their consignees dated June 18, 2010. The letter identified the product, the problem, and the action to be taken. Consignees were instructed to: (1) Discontinue use of all materials provided with I-PTH ELISA kit lot number 992751 and I-PTH IRMA lot numbers 090013, 090121, and 090372. (2) Review historical external quality control results to ensure the assay performance of these lots was within laboratory specifications. Retesting may be necessary if their laboratory uses only the Control materials provided with the I-PTH ELISA and IRMA kits. (3) Abnormal elevated patient results that are obtained with these lot numbers and that are inconsistent with the patient's clinical status should be reviewed in the context of other clinical diagnostic tests. (4) I-PTH ELISA kit lot number 992751 and I-PTH IRMA lot numbers 090013, 090121, and 090372 will no longer be shipped. Due to the calibrator instability recently encountered with DSL-10-8000 I-PTH ELISA and DSL-8000 I-PTH IRMA assays, Beckman Coulter has decided to immediately discontinue these two part numbers. Consignees were asked to share this information with their laboratory staff and retain the notification as part of their laboratory Quality System documentation. Complete and return the enclosed response form within 10 days so Beckman Coulter can be assured that they received this important notification. If they have any unused product the were asked to contact the numbers listed below: (1) In the United States, contact Customer Service at 1-800-526-382.1 (2) Outside of the United States, consignees were asked to contact their local Beckman Coulter Representative Consignees were told to contact DSL Technical Support Center at 1-800-231-7970 in the United States, or their local Beckman Coulter Representative with any questions. For questions regarding this recall call (714) 961-3634.

Distribution

Worldwide Distribution - USA including MD, NY, GA, NJ, CA, MI, TX, PA, IA, and IN. Internationally to Austria, Brazil, China, Colombia, Czech Republic, France, Germany, Hong Kong, India, Japan, Mexico, and the Russian Federation

Quantity

3 kits