FDA Recall Terminated

Symbia T Series SPECT-CT (Single-Photon Emission Computed Tomography and Computed Tomography) System; Siemens Medical Solutions USA, Inc., Hoffman Estates, IL 60195-2061. Indications for use: SPECT + CT: Perform scans and nuclear imaging studies with the same instrument. To obtain attenuation corrected images and to provide registration of anatomical and physiological images within the patient's anatomy.

Recall: Z-0584-2010 · Initiated December 11, 2009

Recall

Recall Number
Z-0584-2010
Event Number
53919
Firm
Siemens Medical Solutions USA, Inc.
FEI Number
1000514697
Product Code
JAK
Status
Terminated
Root Cause
Process control
Initiated
December 11, 2009
Posted
December 29, 2009
Terminated
August 27, 2012
Address
2501 Barrington Rd, Hoffman Estates, IL, 60195-2061

Description

Symbia T Series SPECT-CT (Single-Photon Emission Computed Tomography and Computed Tomography) System; Siemens Medical Solutions USA, Inc., Hoffman Estates, IL 60195-2061. Indications for use: SPECT + CT: Perform scans and nuclear imaging studies with the same instrument. To obtain attenuation corrected images and to provide registration of anatomical and physiological images within the patient's anatomy.

Reason

Some of the hand controllers of the Symbia S & T systems are missing a resistor switch, and the absence of the switch may lead to unintended system motion during camera set up activities.

Action

Siemens issued an Urgent Field Correction Recall letters dated November 18, 2009 and December 11, 2009 to the affected users of the Symbia S and Symbia T Systems, respectivley. The customers were advised of precautionary measures and that Siemens will replace the affected device. They also were advised that a Siemens Customer Service Representative would contact them within a few days to schedule a replacement of the device. For further information, contact Siemens Medical Solutions USA, Inc. at 1-800-888-7436 (USA).

Distribution

Worldwide Distribution -- United States, Algeria, Australia, Austria, Belgium, Canada, Chile, China, Finland, France, Germany, Hong Kong, India, Indonesia, Italy, Korea, Mexico, Netherlands, Norway, Philippines, Portugal, Qatar, Russia, Saudi Arabia, Spain, Sweden, Taiwan, Turkey, Venezuela and Vietnam.

Quantity

97 units