FDA Recall Terminated

Hill-Rom Affinity Four Birthing Bed, Model P3700B. Note: Not distributed in the U.S.

Recall: Z-0584-2008 · Initiated November 21, 2007

Recall

Recall Number
Z-0584-2008
Event Number
45807
Firm
Hill-Rom, Inc.
FEI Number
1824206
Product Code
KNC
Status
Terminated
Root Cause
Device Design
Initiated
November 21, 2007
Posted
February 22, 2008
Terminated
September 22, 2009
Address
125 E Pearl St, Batesville, IN, 47006

Description

Hill-Rom Affinity Four Birthing Bed, Model P3700B. Note: Not distributed in the U.S.

Reason

Hole/opening on the labor bar -- Amputation of a finger could result, if a finger is placed into the opening, as the hole/opening creates a shear point when the mechanism is activated.

Action

Consignees were notified of the problem via an Urgent Medical Device Correction letter dated 11/21/07. Users were informed that they would be receiving a free upgrade kit for all of their Affinity Birthing Beds. The Kit includes two plugs per bed, an insertion tool, and full written instructions. An instruction video is available at www.Hill-Rom.com.

Distribution

Worldwide Distribution-- including USA and countries of Argentina, Australia, Austria, Bahamas, Bahrain, Belgium, Bermuda, Brazil, Brunei, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Dominican Republic, Egypt, Finland, France, Germany, Greece, Guam, Honduras, Hong Kong, Hungary, Iceland, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Lebanon, Malaysia, Mexico, Morocco, The Netherlands, New Calidonia, New Zealand, Nigeria, Northern Mariana Islands, Norway, Oman, Panama, Philippines, Poland, Portugal, Qatar, Republic of Korea, Saudi Arabia, Singapore, South Korea, Romania, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguey, Venezuela, and Virgin Islands .

Quantity

32,379 for all model numbers