Hill-Rom Affinity Three Birthing Bed, Model P3700A.
Recall
- Recall Number
- Z-0583-2008
- Event Number
- 45807
- Firm
- Hill-Rom, Inc.
- FEI Number
- 1824206
- Product Code
- KNC
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- November 21, 2007
- Posted
- February 22, 2008
- Terminated
- September 22, 2009
- Address
- 125 E Pearl St, Batesville, IN, 47006
Description
Hill-Rom Affinity Three Birthing Bed, Model P3700A.
Hole/opening on the labor bar -- Amputation of a finger could result, if a finger is placed into the opening, as the hole/opening creates a shear point when the mechanism is activated.
Consignees were notified of the problem via an Urgent Medical Device Correction letter dated 11/21/07. Users were informed that they would be receiving a free upgrade kit for all of their Affinity Birthing Beds. The Kit includes two plugs per bed, an insertion tool, and full written instructions. An instruction video is available at www.Hill-Rom.com.
Worldwide Distribution-- including USA and countries of Argentina, Australia, Austria, Bahamas, Bahrain, Belgium, Bermuda, Brazil, Brunei, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Dominican Republic, Egypt, Finland, France, Germany, Greece, Guam, Honduras, Hong Kong, Hungary, Iceland, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Lebanon, Malaysia, Mexico, Morocco, The Netherlands, New Calidonia, New Zealand, Nigeria, Northern Mariana Islands, Norway, Oman, Panama, Philippines, Poland, Portugal, Qatar, Republic of Korea, Saudi Arabia, Singapore, South Korea, Romania, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguey, Venezuela, and Virgin Islands .
32,379 for all model numbers