FDA Recall Terminated

Medtronic 2T10R1 Intersept Custom Tubing Pack. Pack 2T10R1 consists of two PVC lines; one of them is 132 inch in length and is capped. The second line is 102 inch in length and has a one way vacuum relief valve assembled in one end. Tubing is used in the vent roller heads of the heart lung machine. This tubing is connected to the cannula in the patient in order to divert blood to the circuit when the patient is on bypass.

Recall: Z-0583-2007 · Initiated January 2, 2007

Recall

Recall Number
Z-0583-2007
Event Number
37079
Firm
Medtronic Perfusion Systems
FEI Number
1000116158
Product Code
DWE
Status
Terminated
Root Cause
Other
Initiated
January 2, 2007
Posted
March 8, 2007
Terminated
February 24, 2008
Address
7611 Northland Dr N, Brooklyn Park, MN, 55428-1088

Description

Medtronic 2T10R1 Intersept Custom Tubing Pack. Pack 2T10R1 consists of two PVC lines; one of them is 132 inch in length and is capped. The second line is 102 inch in length and has a one way vacuum relief valve assembled in one end. Tubing is used in the vent roller heads of the heart lung machine. This tubing is connected to the cannula in the patient in order to divert blood to the circuit when the patient is on bypass.

Reason

One lot of Custom Perfusion Sets, Catalog 2T10R1 was incorrectly assembled. The one way vacuum relief valve on the 102 inch line is attached backwards, thus preventing flow.

Action

There was only one customer. The customer was sent a letter on December 29, 2006, requesting that the impacted product be disposed of or returned to Medtronic for disposal.

Distribution

Illinois

Quantity

17