FDA Recall Terminated

Baxter System 1000 Single Patient Hemodialysis Delivery System; Model SYS1000; all series 1000 instruments labeled as System 1000, AltraTouch 1000, Baxter Tina and Baxter Aurora; Baxter Healthcare Corporation, Deerfield, IL 60015 USA or Baxter Healthcare Corporation, McGaw Park, IL 60085 USA

Recall: Z-0583-04 · Initiated February 20, 2004

Recall

Recall Number
Z-0583-04
Event Number
28375
Firm
Baxter Healthcare Renal Div
FEI Number
3000210122
Product Code
FII
Status
Terminated
Root Cause
Other
Initiated
February 20, 2004
Posted
July 20, 2004
Terminated
February 26, 2008
Address
1620 Waukegan Rd, Bldg R, Mc Gaw Park, IL, 60085-6730

Description

Baxter System 1000 Single Patient Hemodialysis Delivery System; Model SYS1000; all series 1000 instruments labeled as System 1000, AltraTouch 1000, Baxter Tina and Baxter Aurora; Baxter Healthcare Corporation, Deerfield, IL 60015 USA or Baxter Healthcare Corporation, McGaw Park, IL 60085 USA

Reason

These devices have the potential to overheat in the absence of a Ground Fault Circuit Interrupter (GFCI). This heater system failure could result in a fire.

Action

Baxter sent urgent device correction letters dated 2/20/04 to the Hemodialysis Unit Administrators on the same date. The letters informed the accounts of the potential of a heater system failure that under specific conditions could lead to a fire. The letters listed the concurrent conditions that must exist for the potential overheating, and recommended that the following steps be taken to ensure the safety in operating the devices: 1) Plug these machine into GFCI outlets if available, and 2) If GFCI outlets are not available at the facility, the machines must be 'hard-powered off' after dialysis treatments are complete at the end of the day. The machine is had-powered off by placing the main power switch located in the back of the machine into the OFF position. If the facility typically performs a heat clean cycle of machines after treatments are completed, it is imperative that the machines are hard powered off at the completion of the cycle. Any questions were directed to Baxter Instrument Services at 1-800-553-6898, select prompt 3, option 2 and choice 1, Monday through Friday, 8:30 AM to 6:00 PM EST.

Distribution

The devices were distributed nationwide and internationally to Baxter subsidiaries in Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Denmark, Ecuador, Finland, France, Republic of Georgia, Greece, Guatemala, Honduras, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Lebanon, Morocco, Malaysia, the Netherlands, Oman, Panama, Preu, Philippines, Poland, Palestine, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Korea, Spain, Sweden, Taiwan, Thailand, Tunisia, Turkey and United Kingdom.

Quantity

28,482 units