FDA Recall Terminated

Delta/Delta XL Monitors with Omega S Package

Recall: Z-0582-2010 · Initiated November 9, 2009

Recall

Recall Number
Z-0582-2010
Event Number
53875
Firm
Draeger Medical, Inc.
FEI Number
2517967
Product Code
MHX
Status
Terminated
Root Cause
Software design
Initiated
November 9, 2009
Posted
December 29, 2009
Terminated
August 23, 2011
Address
3135 Quarry Rd, Telford, PA, 18969-1042

Description

Delta/Delta XL Monitors with Omega S Package

Reason

monitor is rebooting/resetting

Action

Draeger Medical Systems, Inc. issued and Urgent Medical Device Recall letter dated November 2009 to their affected customers informing them of the problem and a work around until software can be validated. A Technical Service Bulletin has been generated and distributed globally that instructs the affected International Drager subsidiaries to inform customers in their countries. For further information, contact Draeger Medical Systems, Inc. at 1-800-543-5047, extension 1-2349.

Distribution

The product was shipped to medical facilities in CA, CO, CT, ID, IN, KY, MS, NY, OK, and WI. The product was also shipped to United Arab Emirates, Argentina, Austria, Australia, Bosnia Herz, Belgium, Benin, Brunei Darussalam, Bolivia, Brazil, Belarus, Canada, Switzerland, Chile, China, Columbia, Costa Rica, Cyprus, Czech Republic, Germany, Denmark, Egypt, Spain, Finland, France, Great Britain, Greece, Hong Kong, Croatia, Hungary, Ireland, India, Iran, Italy, Japan, Kenya, Korea, Kuwait, Kazakhstan, Lebanon, Lithuania, Latvia, Libya, Morocco, Mexico, Malaysia, Netherlands, Norway, Nepal, Oman, Poland, Portugal, Qatar, Reunion, Romania, Serbia, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovak Republic, Thailand, Turkey, Taiwan, Ukraine, Venezuela, Vietnam, and South Africa.

Quantity

4,188 units