Delta/Delta XL Monitors with Omega S Package
Recall
- Recall Number
- Z-0582-2010
- Event Number
- 53875
- Firm
- Draeger Medical, Inc.
- FEI Number
- 2517967
- Product Code
- MHX
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- November 9, 2009
- Posted
- December 29, 2009
- Terminated
- August 23, 2011
- Address
- 3135 Quarry Rd, Telford, PA, 18969-1042
Description
Delta/Delta XL Monitors with Omega S Package
monitor is rebooting/resetting
Draeger Medical Systems, Inc. issued and Urgent Medical Device Recall letter dated November 2009 to their affected customers informing them of the problem and a work around until software can be validated. A Technical Service Bulletin has been generated and distributed globally that instructs the affected International Drager subsidiaries to inform customers in their countries. For further information, contact Draeger Medical Systems, Inc. at 1-800-543-5047, extension 1-2349.
The product was shipped to medical facilities in CA, CO, CT, ID, IN, KY, MS, NY, OK, and WI. The product was also shipped to United Arab Emirates, Argentina, Austria, Australia, Bosnia Herz, Belgium, Benin, Brunei Darussalam, Bolivia, Brazil, Belarus, Canada, Switzerland, Chile, China, Columbia, Costa Rica, Cyprus, Czech Republic, Germany, Denmark, Egypt, Spain, Finland, France, Great Britain, Greece, Hong Kong, Croatia, Hungary, Ireland, India, Iran, Italy, Japan, Kenya, Korea, Kuwait, Kazakhstan, Lebanon, Lithuania, Latvia, Libya, Morocco, Mexico, Malaysia, Netherlands, Norway, Nepal, Oman, Poland, Portugal, Qatar, Reunion, Romania, Serbia, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovak Republic, Thailand, Turkey, Taiwan, Ukraine, Venezuela, Vietnam, and South Africa.
4,188 units