MICRO-X Rover Mobile X-ray System, # MXU-RV35
Recall
- Recall Number
- Z-0580-2025
- Event Number
- 95624
- Firm
- Micro-X Ltd. Unit 14 6 Mab Eastern Promenade Tonsley Australia
- FEI Number
- 3012855798
- Product Code
- IZL
- Status
- Open, Classified
- Root Cause
- Environmental control
- Initiated
- October 22, 2024
- Posted
- November 27, 2024
Description
MICRO-X Rover Mobile X-ray System, # MXU-RV35
Mobile x-ray system receiving/storage(operation) requirements to be followed: Temp:14 to 131(50 to 86) degrees F, RH:10% to 86%(30% to 60%) allow condensation dry time, Atmospheric Pressure:50kPa to 106 kPa(70kPa to 106kPa), Altitude:-102ft to 12,000 ft, (Maximum Gradient:9 degrees Fahrenheit), which mitigate risk of capacitor malfunction, unpredictable behavior, overheating, unintended start-up.
On 10/22/24, Service Bulletin SB007-23-1.0 was emailed to customers restating receiving, storage, and operation conditions and informing customers that extreme humidity environments can adversely affect the equipment. For further information contact your local service representative: phone: 1-206-249-8764 Email: [email protected]