FDA Recall Terminated

steroEOS Workstation with software versions: 1.3, 1.4 and 1.5 used in processing 3D X-ray images Product Usage: The sterEOS Workstation is a general system for acceptance , transfer, display and digital processing of two 3D x-ray images.

Recall: Z-0580-2014 · Initiated November 29, 2013

Recall

Recall Number
Z-0580-2014
Event Number
66980
FEI Number
3006165260
Product Code
LLZ
Status
Terminated
Root Cause
Software design
Initiated
November 29, 2013
Posted
December 26, 2013
Terminated
January 13, 2015
Address
EOS Imaging 10 rue Mercoeur 4 Ieme Etage Paris France France

Description

steroEOS Workstation with software versions: 1.3, 1.4 and 1.5 used in processing 3D X-ray images Product Usage: The sterEOS Workstation is a general system for acceptance , transfer, display and digital processing of two 3D x-ray images.

Reason

Error copying information to the clinical file and to the report when two 3D models are opened simultaneously on the sterEOS workstation

Action

EOS imaging issued a safety information letter dated November 29, 2013, to the US customer sites. The letter identifed the affected product, problem and actions to be taken. Users of the sterEOS workstation are asked not to simultaneously open several pairs of images in sterEOS 3D. The letter informed customers that EOS imaging will, at it's own expense, corect all systems affected by means of a software update. An EOS imaging representative will contact affected customers to make an appointment for this update once it is available. Users of the sterEOS workstation are asked not to simultaneously open several pairs of images in sterEOS 3D. For questions call +33 (0) 155 25 6060.

Distribution

USA Nationwide Distribution

Quantity

36 units