Invacare Perfecto 2 V Oxygen Concentrator The finished device is packaged as a single unit. The service part contains a single pair of sieve beds; one pair of sieve beds is used in a single finished device. Intended to provide supplemental oxygen to patients with respiratory disorders, by separating nitrogen from room air, by way of a molecular sieve.
Recall
- Recall Number
- Z-0577-2015
- Event Number
- 69441
- Firm
- Invacare Corporation
- FEI Number
- 1525712
- Product Code
- CAW
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- September 23, 2014
- Posted
- December 12, 2014
- Terminated
- June 18, 2015
- Address
- 1200 Taylor St, Elyria, OH, 44035-6248
Description
Invacare Perfecto 2 V Oxygen Concentrator The finished device is packaged as a single unit. The service part contains a single pair of sieve beds; one pair of sieve beds is used in a single finished device. Intended to provide supplemental oxygen to patients with respiratory disorders, by separating nitrogen from room air, by way of a molecular sieve.
It was determined that a defective sieve bed subassembly enabled zeolite sieve dust from the sieve beds to migrate into the valves of the manifold assembly and muffler causing a reduction in oxygen concentration.
Invacare sent an Urgent - Medical Device Recall letter to customers, dated October 2014. The letter identified the affected product as well as the problem involved. The letter instructed customers to examine their device and confirm that it has an affected serial number, or if service was conducted on the device using IRC5O2V Service Part Number 1165099 with the affected date codes. Customers were asked to immediately contact their dealer or service provider at the number listed, so that a replacement can provided. A service provider or dealer should be contacted with questions regarding the instructions.
Nationwide Distribution -- FL, NJ, GA, RI, PA, MI, CA, MA, IN, WA, CT, VA, NC, MO, AZ, CO, MS, MI, WI, TN, NH, IL, LA, OK, HI, KY, NE, SC, TX, IA, IN, and MS.
5,963