FDA Recall Terminated

Stat Profile Critical Care (CCX) Analyzer Catalog Number: 35942

Recall: Z-0576-06 · Initiated January 27, 2006

Recall

Recall Number
Z-0576-06
Event Number
34522
Firm
Nova Biomedical Corporation
FEI Number
1219029
Product Code
CHL
Status
Terminated
Root Cause
Other
Initiated
January 27, 2006
Posted
February 28, 2006
Terminated
October 19, 2006
Address
200 Prospect St, Waltham, MA, 02453-3457

Description

Stat Profile Critical Care (CCX) Analyzer Catalog Number: 35942

Reason

Analyzers may fail to perform scheduled automatic two-point calibrations and Quality Control runs

Action

Nova Biomedical Issued on 1/27/2006 an Advisory Notice to Stat Profile CCX Customers to report the malfunction and request the Good Laboratory Practices (GLP) be used and review be made to assure the calibration status and QC is being performed. A new software version is scheduled to be released to correct the error.

Distribution

Nationwide Foreign: Canada, UAE, Taiwan, Singapore, Budapest, UK, France, Malaysia, France, Mexico, Korea, Finland, Germany, Turkey, Ireland, Italy, Austria, Australia, Netherlands, Venezuela, Japan, Greece, Thailand, Israel, Spain, China, Hong Kong, Argentina, Czech Republic, Croatia,Trinidad WI

Quantity

498 units