FDA Recall Terminated

Access Immunoasay Systems Digoxin Reagent

Recall: Z-0576-05 · Initiated February 23, 2005

Recall

Recall Number
Z-0576-05
Event Number
31253
Firm
Beckman Coulter Inc
FEI Number
2050012
Product Code
KXT
Status
Terminated
Root Cause
Other
Initiated
February 23, 2005
Posted
March 8, 2005
Terminated
September 2, 2005
Address
200 S Kraemer Blvd, Brea, CA, 92822-6208

Description

Access Immunoasay Systems Digoxin Reagent

Reason

Variability in results on Patient and QC results of 20-25%, high and low.

Action

Letters were sent to customers via UPS overnite, they request that alternative methods be used for patient testing until new material is made available. Customers are requested to ask for replacements using a fax back form.

Distribution

Nationwide and Canada

Quantity

4352 domestically.