FDA Recall Terminated

Alinity c Carbon Dioxide Reagent Kit, Produced for Abbott by Sekisui Diagnostics P.E.I. Inc. Prince Edward Island Canada

Recall: Z-0575-2020 · Initiated October 23, 2019

Recall

Recall Number
Z-0575-2020
Event Number
84140
Firm
Abbott GmBH & Co. KG Max-Planck-Ring 2 65205 Wiesbaden Germany
FEI Number
3002809144
Product Code
KHS
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
October 23, 2019
Terminated
May 14, 2021

Description

Alinity c Carbon Dioxide Reagent Kit, Produced for Abbott by Sekisui Diagnostics P.E.I. Inc. Prince Edward Island Canada

Reason

Abbott internal studies have determined that atmospheric Carbon Dioxide (CO2) can be absorbed into the Alinity c Carbon Dioxide reagent, resulting in the potential for incorrect results. Abbott was made aware of this issue from customer complaints. Internal testing demonstrated that the amount of CO2 absorbed is higher with increased reagent carousel rotation and when the volume of reagent in the cartridge is reduced. This phenomenon can be detected as a shift in QC. The shift is predictive in both direction and magnitude of the potential impact to patient results.

Action

On 10/23/19, a Product Recall letter was issued to all worldwide Alinity c Carbon Dioxide Reagent customers who received the in-date lots. US customers were sent this information via FedEx priority overnight. Customers are asked to immediately discontinue use and destroy product from LN 07P7230 lot 54449UQ03. For LN 07P7220 lot 54454UQ03, the product may continue to be used but 2 levels of CO2 controls must be run every hour until the new material is received. If you or any of the health care providers you serve have any questions regarding this information, US Customers can contact Customer Service at 1-877-4ABBOTT (available 24 hours a day, 7 days a week). Customers outside the US , please contact your local area Customer Service.

Distribution

Worldwide and US Nationwide in the states of: AL, AR, CA, CO, FL, GA, ID, IL, LA, MA, MI, MN, MO, NC, NY, OK, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, and Puerto Rico Foreign: AUSTRALIA, BANGLADESH, BELGIUM, BOTSWANA, BRAZIL, CANADA, COLOMBIA, DENMARK, FRANCE, GREECE, Hong Kong, INDIA, ITALY, JAMAICA, LEBANON, MAURITIUS, NAMIBIA, NETHERLANDS, NEW ZEALAND, PAKISTAN, PHILIPPINES, POLAND, PORTUGAL, PORTUGAL, ROMANIA, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TRINIDAD & TOBAGO, UAE, and UNITED KINGDOM

Quantity

105 kits (lot 54449UQ03) and 1215 kits (lot 54454UQ03)