FDA Recall
Terminated
StarClose Vascular Closure System (Clip Applier & Exchange System), Catalog Number: 14677; Product is manufactured by: Abbott Vascular Inc., 400 Saginaw Drive, Redwood City, CA 94063
Recall: Z-0575-2007
·
Initiated January 31, 2007
Recall
- Recall Number
- Z-0575-2007
- Event Number
- 37270
- Firm
- Abbott Vascular, Inc.
- FEI Number
- 3000718335
- Product Code
- DYB
- Status
- Terminated
- Root Cause
- Other
- Initiated
- January 31, 2007
- Posted
- March 8, 2007
- Terminated
- February 26, 2008
- Address
- 400 Saginaw Dr, Redwood City, CA, 94063-4749
Description
StarClose Vascular Closure System (Clip Applier & Exchange System), Catalog Number: 14677; Product is manufactured by: Abbott Vascular Inc., 400 Saginaw Drive, Redwood City, CA 94063
Reason
Premature release of the vessel locator wings, which stabilize the device prior to clip deployment, will result in no hemostasis.
Action
On 1/31/07, the firm initiated the recall and its notification was via letters explaining the reason for the recall and requesting the product be returned to its sales rep.
Distribution
Worldwide distribution, including USA, Puerto Rico, and 410 other countries. There were no shipments to Canada and Mexico.
Quantity
97,952 units