FDA Recall Terminated

ARCHITECT HAVAB-G Calibrator, REF 6L27-01

Recall: Z-0574-2020 · Initiated November 7, 2019

Recall

Recall Number
Z-0574-2020
Event Number
84238
Firm
Abbott Gmbh & Co. KG Max-Planck-Ring 2 65205 Wiesbaden Germany
FEI Number
3002809144
Product Code
LOL
Status
Terminated
Root Cause
Material/Component Contamination
Initiated
November 7, 2019
Posted
November 26, 2019
Terminated
August 31, 2023

Description

ARCHITECT HAVAB-G Calibrator, REF 6L27-01

Reason

Abbott has identified that Calibrator lot 08219BE00 is potentially impacted by an increase in Relative Light Units (RLUs) up to 94%. It has been confirmed that the increase in the calibrator RLU values may cause a decrease in the control values and patient results. Patient values generated using the impacted calibrator lot 08219BE00 may have changed interpretation from reactive to falsely non reactive. Results impacted may present as a (falsely) non reactive result that fall in the range between 0.50 to 0.99 S/CO.

Action

The firm initiated the recall by letter on 11/07/2019. The letter explained the issue and requested that the consignee discontinue use of the product and destroy it. The firm is providing replacement material.

Distribution

US in the states of: AZ, CA, CO, CT, FL, KY, LA, MA, MD, MI, MN, MO, NJ, NY, OH, OR, RI, SC, TN, TX, WA, WV

Quantity

83 kits