AB5000 Console Circulatory Support System Catalog Number: 0015-0000
Recall: Z-0574-2007
·
Initiated January 8, 2007
Recall
- Recall Number
- Z-0574-2007
- Event Number
- 37170
- Firm
- Abiomed, Inc.
- FEI Number
- 1220648
- Product Code
- DSQ
- Status
- Terminated
- Root Cause
- Other
- Initiated
- January 8, 2007
- Posted
- March 7, 2007
- Terminated
- February 7, 2008
- Address
- 22 Cherry Hill Dr, Danvers, MA, 01923-2575
Description
AB5000 Console Circulatory Support System Catalog Number: 0015-0000
Reason
Unit may alarm ''Low Flow'' and ''Low Pressure'' due to manufacturing material in the pressure source.
Action
Abiomed notified customers by letter dated 1/08/07. Service reps will inspect units and replace the defective components.
Distribution
Worldwide, including USA, Japan, Germany, Australia, and Mexico.
Quantity
55 units