Accu-Chek Inform blood glucose test strips; Reference numbers 04535057001, 04647696001 and 04535120001 (All sold outside of the United States).
Recall
- Recall Number
- Z-0573-06
- Event Number
- 34485
- Firm
- Roche Diagnostics Corp.
- FEI Number
- 1823260
- Product Code
- LFR
- Status
- Terminated
- Root Cause
- Other
- Initiated
- February 10, 2006
- Posted
- February 28, 2006
- Terminated
- February 8, 2007
- Address
- 9115 Hague Rd, Indianapolis, IN, 46256-1025
Description
Accu-Chek Inform blood glucose test strips; Reference numbers 04535057001, 04647696001 and 04535120001 (All sold outside of the United States).
There may be a thin piece of plastic across the mouth of the vial, preventing the cap from sitting snugly onto the vial and, if present, this situation will cause erroneously low or high blood glucose readings..
Notifications (letters) dated 2/10/06 were sent to healthcare professionals, retail pharmacies, hospitals and mail order firms asking them to provide their customers with information instructing the users to inspect their strip vials and to not use strips from defective vials. Mail order accounts were instructed to forward the recall notification to end users.
Nationwide, Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Ecuador, England, France, Germany, Hong Kong, India, Indonesia, Japan, Malaysia, Mexico, New Zealand, Pakistan, Panama, Peru, Philippines, Poland, Russia, Singapore, South Africa, South Korea, Spain, Sweden, Taiwan and Thailand.