FDA Recall Terminated

Instrument Manager driver : Driver sysu100i that connects to the Sysmex UF-100 analyzer Software all versions prior to( v7.00.0006 and v8.00.0005): v7.00.0001, v7.00.0002, v7.00.0003, v7.00.0004, v7.00.0005, v8 .00.0001, v8.00.0002, v8.00.0003, and v8 .00.0004

Recall: Z-0571-2014 · Initiated November 6, 2013

Recall

Recall Number
Z-0571-2014
Event Number
66754
Firm
Data Innovations, Inc.
FEI Number
3001451449
Product Code
JQP
Status
Terminated
Root Cause
Software design
Initiated
November 6, 2013
Posted
December 24, 2013
Terminated
December 5, 2014
Address
120 Kimball Ave, Ste 100, South Burlington, VT, 05403-6837

Description

Instrument Manager driver : Driver sysu100i that connects to the Sysmex UF-100 analyzer Software all versions prior to( v7.00.0006 and v8.00.0005): v7.00.0001, v7.00.0002, v7.00.0003, v7.00.0004, v7.00.0005, v8 .00.0001, v8.00.0002, v8.00.0003, and v8 .00.0004

Reason

When connected to the Sysmex UF-100, the Instrument Manager driver sysu100i driver software versions 7.00.0001, 7.00.0002, 7.00.0003, 7.00.0004, 7.00.0005, 8.00.0001, 8.00.0002, 8.00.0003, and 8.00.0004, and is configured with option "Strip Zeros from Result" set to "Yes" may result in sample error

Action

Data Innovation notified consignees by Important Medical Device Recall Notification dated 11/5/13 to all list serv recipients of the problem and posted critical alert on Customer Web Portal. A certified letter was issued to impacted customers and consignees if no response was recieved from e-mail.

Distribution

Nationwide Canada and Britain

Quantity

24 systems