FDA Recall Terminated

Ventana Image Analysis System (VIAS) - PATHWAY anti-HER-2/neu (4B5) primary antibody with software version 2.3.0.2, Model Number: VIAS-799-10000-00, Pathology Image Analysis System, Ventana Medical Systems, Inc., Tucson, AZ 85755

Recall: Z-0571-2008 · Initiated June 26, 2007

Recall

Recall Number
Z-0571-2008
Event Number
45798
Firm
Ventana Medical Systems Inc
FEI Number
1000125183
Product Code
NQN
Status
Terminated
Root Cause
Software change control
Initiated
June 26, 2007
Posted
March 18, 2008
Terminated
March 26, 2012
Address
1910 E Innovation Park Dr, Tucson, AZ, 85755-1962

Description

Ventana Image Analysis System (VIAS) - PATHWAY anti-HER-2/neu (4B5) primary antibody with software version 2.3.0.2, Model Number: VIAS-799-10000-00, Pathology Image Analysis System, Ventana Medical Systems, Inc., Tucson, AZ 85755

Reason

Incomplete upgrade: The software upgrade from February 1, 2007 may not have been completed. This software upgrade adds the algorithm (Slidetype) for the PATHWAY anti-HER-2/neu (4B5) antibody to the VIAS system.

Action

Beginning in 6/07, Ventana Medical Systems, Inc., made phonecalls to labs asking to confirm software version, date of upgrade if available and confirmation of whether site uses 4B5 and Imaging together. A phone script was provided. Customers with systems not ugraded were to upgrade before further use of system. A notification letter dated 8/17/2007 was later sent to VIAS customers advising that the software upgrade from February 1 , 2007 may not have been completed on their system. This software upgrade adds the algorithm (Slidetype) for the PATWAY anti-HER-2/neu (4B5) antibody to their VIAS system . This Slidetype is cleared for marketing and is required prior to imaging slides stained with PATHWAY anti-HER-2/neu (4B5). The letter also advises that their VIAS and NEXES systems are conneded to the internet which enables this upgrade to be performed remotely with the support of the Ventana Medical Systems Customer Support Center (CSC). customers are instructed to call the CSC to carry out the upgrade or to schedule a Technical Applications Specialist (TAS) to assist with the upgrade in their lab. Arrangements for the upgrade are to be made prior to imaging slides stained with PATHWAY anti-HER-2/neu (4B5).

Distribution

Nationwide and Canada

Quantity

77 devices