Brand Name: Single Use Ligating Device Product Name: Olympus HX-400U-30 Model/Catalog Number: HX-400U-30 Used with an Olympus endoscope to deliver a nylon loop snare.
Recall
- Recall Number
- Z-0570-2026
- Event Number
- 97731
- Firm
- Olympus Corporation of the Americas
- FEI Number
- 2429304
- Product Code
- MND
- Status
- Open, Classified
- Root Cause
- Process control
- Initiated
- October 30, 2025
- Posted
- November 25, 2025
- Address
- 3500 Corporate Pkwy, Center Valley, PA, 18034-8229
Description
Brand Name: Single Use Ligating Device Product Name: Olympus HX-400U-30 Model/Catalog Number: HX-400U-30 Used with an Olympus endoscope to deliver a nylon loop snare.
Potential that the ligation loop was unable to release or detach as expected during use, causing the loop to become unintentionally fixed around patient anatomy.
On October 30, 2025 and again on November 3, 2025, URGENT: MEDICAL DEVICE CORRECTIVE ACTION letters were sent to customers. Actions to be taken: Users are reminded of the importance of adhering to the Warnings and Cautions statements that are present in the instructions for use (IFU, ref: GK4574 Rev 16). A diagram was provided to illustrate the need to maintain a straight position of the proximal end of the tube sheath to prevent increased frictional resistance which may lead to breakage of the operating piipe when the slider is pushed. Particular attention should be paid to the following key guidance within Sections 10 and 12 of the IFU. Carefully read the content of this letter, including the supplemental guidance and reinforced text from the IFU as noted above. a) Always have Olympus loop cutter available when using the PolyLoop device b) When operating the yellow tube joint (yellow cylinder), be careful not to move it away from the handle, prematurely tightening the loop which may cause the inability to release the ligation loop. c) If resistance is encountered, do not continue or force the advancement of the slider located on the device handle. 2. Ensure all personnel are completely knowledgeable on the content of this notification. You may continue to use the device and are reminded of the importance of adhering to the warnings that are present in the instructions for use. 3. Olympus requests that you acknowledge receipt of this letter through our recall web portal. 4. If you have further distributed this product, identify your customers, and forward this notification to them. ***Update 04/24/2026*** On April 20, 2026 URGENT: MEDICAL DEVICE CORRECTIVE ACTION UPDATE letters were sent to customers. Olympus has updated the Instructions for Use (IFU) for the Single Use Ligating Device to clarify correct handling and loop-release techniques. The revised IFU includes previously issued guidance as well as supplemental explanations and illustrations intended to clari
US-wide distribution
7803 units