FDA Recall Open, Classified

Biolox Ceramic Option Head Zimmer 12/14 40mm +0, Item Number 00-8777-040-02

Recall: Z-0570-2024 · Initiated November 2, 2023

Recall

Recall Number
Z-0570-2024
Event Number
93486
Firm
Biomet, Inc.
FEI Number
1825034
Product Code
LZO
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
November 2, 2023
Posted
December 18, 2023
Address
56 E Bell Dr, Warsaw, IN, 46582-6989

Description

Biolox Ceramic Option Head Zimmer 12/14 40mm +0, Item Number 00-8777-040-02

Reason

One product complaint was received reporting that there was an incorrect adapter in the packaging. An adapter with neck length M/+0 was inside the product packaging, which should have contained an adapter with neck length S/-3.0. The difference in neck length may be recognized by the size (S or M) indicator on the device.

Action

An URGENT MEDICAL DEVICE RECALL notification letter dated 11/2/23 was sent to customers. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the recall notice for hospitals and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected]. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 3. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Frida

Distribution

US States: AL, CA, MD, MI, NC, OH, OK, TN, TX. Netherlands.

Quantity

10 units