Scissor Tips Model #'s: AESPO888 AESPO889 ENCES0102 MIC3112 MIC3122 MIC3142 MIC3152 SNO89-5100 and SNO89-5300 Intended to be used with a resusable hand piece and are designed to use in minimally invasive and open surgical procedures to facilitate coagulation, preparation, mobilization, and cutting tissue.
Recall
- Recall Number
- Z-0570-2009
- Event Number
- 49510
- Firm
- Sterilmed Inc
- FEI Number
- 3001237663
- Product Code
- NUJ
- Status
- Terminated
- Root Cause
- Packaging process control
- Initiated
- August 8, 2008
- Posted
- January 13, 2009
- Terminated
- April 26, 2011
- Address
- 11400 73rd Ave N, Ste 100, Osseo, MN, 55369-5562
Description
Scissor Tips Model #'s: AESPO888 AESPO889 ENCES0102 MIC3112 MIC3122 MIC3142 MIC3152 SNO89-5100 and SNO89-5300 Intended to be used with a resusable hand piece and are designed to use in minimally invasive and open surgical procedures to facilitate coagulation, preparation, mobilization, and cutting tissue.
SteriMed, Inc is conducting a product removal of a small group of devices individually packaged and sealed within a pouch. The removal is being conducted because the possibility exists that some of the pouches were not properly sealed on one end. A breach in packaging seal or a failure in packaging integrity has the possibility of risk to the patient in terms of transmitting organisms capable
Consignees were notified of the recall via verbal communication in person by on-site technician and or their local sales representative on 8/8/08. Others were contacted via fax or e-mail followed by on-site visits 8/11/08. SteriMed also provided consignees (Materials Manager/Risk Manager) with a "Urgent Product Removal" recall letter (August 14, 2008) in person informing them of the potential of receiving affected product. Included in the Product Removal Notification was a Customer Reply Form to inventory quantity of product received and quantity returned. Contact SterilMed, Inc. at 1-888-541-0078 for assistance.
AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, MA, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, WA, WI, and WV OUS: CANADA
161