BD Biosciences MultiSET Flow Cytometry Software Versions 2.1, 2.1.1, and 2.2 with OS X operating systems, BD Biosciences, San Jose, CA 95131.

Recall: Z-0570-2008 · Initiated September 6, 2007

Recall

Recall Number
Z-0570-2008
Event Number
45447
Firm
BD Biosciences
FEI Number
2916837
Product Code
GKZ
Status
Terminated
Root Cause
Software design
Initiated
September 6, 2007
Posted
March 4, 2008
Terminated
April 15, 2008
Address
2350 Qume Dr, San Jose, CA, 95131-1807

Description

BD Biosciences MultiSET Flow Cytometry Software Versions 2.1, 2.1.1, and 2.2 with OS X operating systems, BD Biosciences, San Jose, CA 95131.

Reason

Inaccurate results: Software error results in inaccurate display result statistics. If the user adjusts the lymph gate or attractors in the Lab Report view without using the Manual Gate function (as described in the system manual), the statistic results will not be updated to match the adjusted gate.

Action

Becton Dickinson sent an Important Product Information letter by mail on 09/14/2007 informing customers of the need to follow the instructions for adjusting the lymph gate or attractors. The next software revision will alleviate this limitation.

Distribution

Worldwide Distribution - USA, Belgium, Brazil, Peru, Singapore, Canada, Jamaica, Chile, Ecuador, Hong Kong-China, Thailand, Mexico, Venezuela, Barbados, Philippines, Puerto Rico, Guatemala, Japan, Uruguay, Argentina, China, Trinidad and Tobago, Honduras, Australia, New Zealand, and India.

Quantity

1136 units