FDA Recall Terminated

Accu-Chek Advantage II blood glucose test strips; Reference numbers 12030659001, 12030667001, 03137872001, 03137899001, 03000281001, 12030535001, 12030543001, 03033449001, 03033465001 and 12030543047 (All sold outside of the United States).

Recall: Z-0568-06 · Initiated February 10, 2006

Recall

Recall Number
Z-0568-06
Event Number
34485
Firm
Roche Diagnostics Corp.
FEI Number
1823260
Product Code
LFR
Status
Terminated
Root Cause
Other
Initiated
February 10, 2006
Posted
February 28, 2006
Terminated
February 8, 2007
Address
9115 Hague Rd, Indianapolis, IN, 46256-1025

Description

Accu-Chek Advantage II blood glucose test strips; Reference numbers 12030659001, 12030667001, 03137872001, 03137899001, 03000281001, 12030535001, 12030543001, 03033449001, 03033465001 and 12030543047 (All sold outside of the United States).

Reason

There may be a thin piece of plastic across the mouth of the vial, preventing the cap from sitting snugly onto the vial and, if present, this situation will cause erroneously low or high blood glucose readings.

Action

Notifications (letters) dated 2/10/06 were sent to healthcare professionals, retail pharmacies, hospitals and mail order firms asking them to provide their customers with information instructing the users to inspect their strip vials and to not use strips from defective vials. Mail order accounts were instructed to forward the recall notification to end users.

Distribution

Nationwide, Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Ecuador, England, France, Germany, Hong Kong, India, Indonesia, Japan, Malaysia, Mexico, New Zealand, Pakistan, Panama, Peru, Philippines, Poland, Russia, Singapore, South Africa, South Korea, Spain, Sweden, Taiwan and Thailand.