FDA Recall Open, Classified

Microsurgical Operating Room Tables;7300BX (18-351-70).

Recall: Z-0567-2021 · Initiated September 11, 2019

Recall

Recall Number
Z-0567-2021
Event Number
85442
Firm
Mizuho America, Inc.
FEI Number
1000159388
Product Code
GDC
Status
Open, Classified
Root Cause
Device Design
Initiated
September 11, 2019
Address
30057 Ahern Ave, Union City, CA, 94587-1234

Description

Microsurgical Operating Room Tables;7300BX (18-351-70).

Reason

Unintended tilting of the Tabletop MST 7300BX and incorrect operation due to infrared operating signal errors of Tabletops MST 7300B and MST 7300BX.

Action

The firm's representatives notified their U.S. consignees regarding the recall notification by phone and a separate FedEx letter was sent (08/09/2019 - date the letter was issued). The Customer Notification Letter "IMPORTANT VOLUNTARY RECALL NOTIFICATION", informed customers of the two (2) types of errors: 1) Unintended Tilting of the Tabletop - Impurities in the hydraulic oil has caused minor leaking inside a control valve causing the table to tilt after remaining static for long periods of time. 2) Incorrect Operation due to Infrared Operating Signal Errors - The table can inadvertently mix infrared signals from the hand pendant while using in the wireless mode and may cause unintended movement. Customer are also being informed that a sales representative will contact them to organize the logistics of the correction for the tables. The recall notification includes to visit the firm's webpage (www.mizuho.com) to find additional information related to the recall. For additional questions, the Customer Service Group or local sales representatives can be contacted.

Distribution

US: AZ, CA, FL, MO, NC, OH, PA, TX, and WI. OUS: Greece, Israel, Japan, and Ukraine.

Quantity

10